XLIMUS Technical Data

XLIMUS Technical Data
 Material  Cobalt Chromium Alloy L-605
 Strut thickness  73μm - 0.073mm - 0.0029“
 Coating Layer  2µm
 Device lengths  Stent length = balloon length = markers distance
 Metal to artery ratio  14% average
 Nominal pressure  8 ATM
 Rated burst pressure  16 ATM except diameters 4.5 / 5.0 and diameter 4.0 with length higher than 20mm (14 ATM)
 Average foreshortening  < 1%
 Guiding catheter compatibility  5F (0.058“ ID) except diameters 4.5 and 5.0 —> 6F (0.071“)
 Guidewire compatibility  0.014“ maximum recommended
 Sirolimus (Rapamycin)
 Drug-Coating
 1.25µg/mm² stent surface