XLIMUS®

Sirolimus Eluting
Coronary Stent System

XLIMUS Technical Data
Material Cobalt Chromium Alloy L-605
Strut thickness 73μm - 0.073mm - 0.0029“
Coating Layer 2µm
Device lengths Stent length = balloon length = markers distance
Metal to artery ratio 14% average
Nominal pressure 8 ATM
Rated burst pressure 16 ATM except diameters 4.5 / 5.0 and diameter 4.0 with length higher than 20mm (14 ATM)
Average foreshortening < 1%
Guiding catheter compatibility 5F (0.058“ ID) except diameters 4.5 and 5.0 —> 6F (0.071“)
Guidewire compatibility 0.014“ maximum recommended
Sirolimus (Rapamycin)
Drug-Coating
1.25µg/mm² stent surface